Pressure Transducer |
||
|
||
Features
|
![]() |
|
| . | ||
| Performance Specifications @ 72°F (22°C) | ||
| Linearity & Hysteresis (Terminal Based Linearity) | < ±
0.25% of span < ± 0.10% of span specify option S10 (Consult factory for ranges) |
| Repeatability | Within ± 0.08% of calibrated span |
| Resolution | Infinite |
| Operating Temperature Range | -13° to 185°F, (-25° to 85°C) |
| Compensated Temperature Range | 0° to 160°F, (-18° to 71°C) |
| Thermal Zero Shift | < ± 0.0075% F.S.O./°F (< ± 0.0135% F.S.O./°C) |
| Thermal Span Shift | < ± 0.0075% F.S.O./°F (< ± 0.0135% F.S.O./°C) |
| Response Time | 100 to 325 microseconds, range dependant, (Consult factory for specific data) |
| Electrical Specifications |
| Bridge | Four active arm bonded strain gage |
| Output | |
| Load Resistance | 2500 Ohms min. |
| Load Capacitance | 0.01 µF max. |
| Supply Voltage | 15 to 30 VDC, unregulated |
| Power Supply Effect | < ± 0.01% of span between 16-32 VDC |
| Insulation Resistance | 1000 Megohms min. @ 50 VDC to transmitter case |
| Supply Current (Quiescent) | 24 mA DC typical |
| Circuitry | |
| Noise | 3 mV peak to peak |
| Output Impedance | 1 ohm max. |
| Circuit Protection | Reverse polarity protection |
| Voltage transient & Surge | Protected per Mil-STD-704 |
| Zero Adjustment | ± 5% of span |
| Span Adjustment | ± 2% of span |
| Shunt Calibration | 80% ± 5% of span (specific value noted on calibration sheet) |
| Mechanical Specifications Dimensional Outline Drawing |
| Pressure Ranges | 50, 100, 250, 500, 1k, 1.5k, 2k, 3k, 5k, 7.5k, 10k, 15k, 20k, 30k, 40k, 50k, 60k | |
| Proof Pressure | 150% F.S. without exceeding specified tolerance | |
| Burst Pressure (Burst pressure is limited to process connection) | > 3 times F.S. or 150k PSI, whichever is less | |
| Wetted Material | A286, 15-5 PH body and 316 S.S. process connection | |
| Option S2 | A286 body and 316 S.S. process connection, Hydrogen and sour gas compatible meets NACE MR0175 latest revision | |
| Weight | 25 ounces approximately | |
| Process Connections | ||
| 50 thru 10,000 PSI | 1/4-18 NPT
Male MS-33649-04 |
|
| 500 thru 60,000 PSI | F250C | |
| Specifications are subject to change without notice. Transducer is supplied with NIST traceable certificate of conformance. 13 point calibration certificate is also available. |